D
5180483-100, R
IRECTION
7-8-5
Trackball
The trackball is low sensitive!
Run Diagnostics for the Op Panel.
PROPRIETARY TO GE
. 7
EV
Figure 7-148 Trackball Troubleshooting
Does the trackball
move at all?
NO
NO
Replace the trackball unit.
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 P
YES
Check Speed and Acceleration
menu: System, General Presets.
S
M
ROPRIETARY
ERVICE
ANUAL
7 - 183