Log Analysis - R4.x (cont'd)
3.) Power Supply Errors
; Info ; acqHouseKeeping(3016); TxPs Alarm source[System (rack) supply] type[No alarm] detail[TS On Strobe]
; Info ; acqHouseKeeping(3020); TxPs alarm status AC-V[118.6 VAC] AC-Freq[60 Hz] ACErr[no] RackOn[48V LV ACinv]
; Info ; acqHouseKeeping(3020);
+12V[11.01 Volts] +6V[5.90 Volts] -6V[-5.99 Volts] +5V[4.99 Volts]
; Info ; acqHouseKeeping(3024); TxPs alarm info source[Transmit supply 2] V+[0.32 Volts] V-[0.00 Volts] Power[0.00 Watts]
Current[0.000 Amps] Prgmd V[0.00 Volts]
; Info ; acqHouseKeeping(3020);
; EScan.View(2004); ESStatus(0,Live): SetState RunState("Freeze" -> "Live") Modes("2D (B-Mode)") Probe("fla_M6-15L")
Appl("MscSkel")
; acqGfi(3860); scanctrl:ScanControl::startScan: still waiting for TS_LEVEL_OK after 200 ms
; acqGfi(3860); scanctrl:ScanControl::startScan: still waiting for TS_LEVEL_OK after 400 ms
; acqGfi(3860); scanctrl:ScanControl::startScan: still waiting for TS_LEVEL_OK after 600 ms
; acqGfi(3860); scanctrl:ScanControl::startScan: still waiting for TS_LEVEL_OK after 800 ms
; acqGfi(3860); scanctrl:ScanControl::startScan: still waiting for TS_LEVEL_OK after 1000 ms
; DFEManager(3900); HandleRunFlag: setting RunBlock flag to false!
; ScanGeneric(2004); SetParam: scanID=1, paramId=23, LookupPort returned NULL!
; EScan.Main(2004); GetParam("RunState") should be replaced by GetParam("InLive")!
; EAErrorHandler(3900); Stops scan
; EAErrorHandler(3900); Error from HouseKeeping: GFI Processing Exception: SubType 101: scanctrl ScanControl::startScan:
Tx Power Supply is not ready. TS_LEVEL_OK ti m eout after 1002 ms, ScanSequencer NOT started! Hardware Status:-GFE Ch0
empty-GFE Ch1 empty-GTX1 HVFAULT-TS_OK_N-TS_LEVEL_OK_N-ACFAIL-THOR_PROBE_ERROR
; acqHouseKeeping(2004); CHouseKeeping::HandleGfiErrorEvents: Gfi Error Event Received: GFI Processing Exception:
SubType 101: scanctrl ScanControl::startScan: Tx Power Supply is not ready. TS_LEVEL_OK timeout after 1002 ms,
ScanSequencer NOT started! Hardware Status:-GFE Ch0 empty-GFE Ch1 empty-GTX1 HVFAULT-TS_OK_N-
TS_LEVEL_OK_N-ACFAIL-THOR_PROBE_ERROR
PROPRIETARY TO GE
. 7
EV
+48V[48.02 Volts] +24V[24.01 Volts] +15V[15.03 Volts] -15V[-15.08 Volts]
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 P
inTemp[38 °C] outTemp[42 °C] +PMX V[0.00 Volts] -PMX V[0.00 Volts]
S
M
ROPRIETARY
ERVICE
ANUAL
7 - 249