Pop-Up: Application Has Performed An Unexpected Halt - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
7-9-18-3
Key Log Files (cont'd)
7-9-18-5-3
LCD Suddenly Goes Black
Condition: System is performing emergency shutdown due to loss of power
Action: Check log\logfile.txt for message about battery operation
Ensure operator waits for LED to go off before trying to turn system on
Example:
Error
Info
Info
Info
process
7-9-18-5-4

POP-UP: Application has Performed an Unexpected Halt

Condition: Application Software Crash
Action: Check for files log\CrashDumps\CrashDump_*.dmp
Correction: Not usable in field – refer to engineering for diagnosis
Example:
Figure 7-174
7 - 266
PROPRIETARY TO GE
. 7
EV
; acqTransLayer(316); AC Failure detected - System will shut down
; acqHouseKeeping(192); CHousekeeping::HandleAcFailEvent() called
; acqHouseKeeping(192); Signaling event "SYSTEM_ON_BATTERY_BACKUP"
; EchoSysmon:BatteryPower(3548); Battery Event signaled, beginning shutdown
Section 7-9 - Troubleshooting Tips
LOGIQ E9 P
ROPRIETARY
S
M
ERVICE
ANUAL

Advertisement

Table of Contents
loading

Table of Contents