LOGIQ E9 VPD Editor "Not Opening"

If a LOGIQ E9 console model running Application Software R3.1.0 and the VPD Editor does not open,
or takes too long to open, the VPD Editor, stand alone times out too quickly before it is able to establish
communication with the Card Rack.
If the system application is running, the VPD Editor will open immediately since the communication to
the Card Rack is already established by the system application.
This information s intended to inform the Field about a work around in case this issue is observed.
Table 7-99
Insert a Service Key:
• Exit application and go to Windows
desktop.
• Double-click on the "Vpdedit" (VPD
Editor) icon, on the desktop and wait.
If the VPD Editor does not open within
the next 30 seconds:
Double-click on the "Go Ichiro" icon to
launch the system application and keep
access to Windows.
NOTE: If the application does not start,
power-cycle the system (full power-
cycle, not re-start) and start procedure
over.