Wiring

13.3.2 Connecting the connecting cable to sensor

Use the cable end without shield wire to connect to the sensor (observe labeling).
NOTICE
If the cable is not terminated or was shortened, observe the separate instructions in the
NOTICE
included terminating set.
1. Loosen the four screws from the neck cover and remove cover.
2. Remove connector nut from cable gland and pull out clamped insert.
3. Push cable through connector nut and clamped insert.
4. Remove precut outer casing of cable.
5. Pull back outer shield of cable over clamped insert.
6. Feed cable through cable entry into sensor.
7. Use the operating tool to connect wires to connection terminals in accordance with
8. Assemble the cable gland and tighten connector nut.
9. Place cover onto sensor and fasten with four screws.
124 / 194
terminal diagram, see Connection terminals [} 121].
A5E52748375-AA, 2023/09
Connecting cable installation