Sony FVS-1000 Maintenance Manual page 21

Multi resolution film scanner
Hide thumbs Also See for FVS-1000:
Table of Contents

Advertisement

4. Remove the two screws, then remove the connector cover from the rear panel.
5. Pass the connector of the air blow assembly through the hole of the connector cover, then connect it
with the connector fixed to the connector support plate.
6. Fix the connector cover to the rear panel with the two screws (B3 x 6).
n
Put the harness inside as much as possible.
Air blow assembly
7. Insert the tube (ø8) to the elbow (one-touch joint) of the air blow assembly.
n
Insert the tube completely, then confirm that it does not come out by pulling slightly.
8. Insert the connecting elbow to the bulkhead union.
n
Insert the connecting elbow completely, then confirm that it does not come out by pulling slightly.
Air blow assembly
FVS-1000 MMP1
Connector
Bulkhead union
Elbow
B3 x 6
Connector cover
Connecting elbow
assembly
Tube (ø8)
1-10. Installation Procedure