Acceptance Criteria Nc (Normal Condition); Acceptance Criteria Sfc (Single Fault Condition) - Ground (Earth), Line Or Neutral Open7 - GE B30 Technical Reference Manual

Patient monitor
Hide thumbs Also See for B30:
Table of Contents

Advertisement

(5)
Read and record the current leakage indicated on the tester.
(6)
Change leakage tester switches to:
(7)
Read and record the current leakage indicated on the tester.
(8)
Change leakage tester switches to:
(9)
Read and record the current leakage indicated on the tester.
If measured reading is greater than the appropriate specification below, the device under test
fails. Contact GE Technical Support.

2.7.1 Acceptance criteria NC (Normal condition)

(USA only) 300 µA, and the device under test is powered from 100-120 V/50-60 Hz
(USA only) 300 µA, and the device under test is powered from a center-tapped 200-240
V/50-60 Hz, single phase circuit (UL Split Phase Exemption)
500 µA, and the device under test is powered from a non-centertapped, 200-240 V/50-60
Hz, single-phase circuit
2.7.2 Acceptance criteria SFC (Single fault condition) - ground (earth), line or
neutral open
(USA only) 300 µA, and the device under test is powered from 100-120 V/ 50-60 Hz
(USA only) 300 µA, and the device under test is powered from a center-tapped 200-240
V/50-60 Hz, single phase circuit (UL Split Phase Exemption)
1000 µA
NOTE: Center-tapped and non-center-tapped supply circuits produce different leakage
currents and the UL and IEC limits are different.
Neutral - OPEN
GND (Earth) - CLOSED
Polarity - NORMAL
Neutral - OPEN
GND (Earth) - CLOSED
Polarity - NORMAL
Neutral - OPEN
GND (Earth) - CLOSED
Electrical Safety Tests
3-7
Document no. 2044677-001