B30 Patient Monitor
6-26
.
Document no
2044677-001
-
Remove the four screws (T6) with washers holding the
NIBP pump to the frame.
-
Carefully remove the front unit together with the NIBP
pump.
-
Disconnect the hoses from the manifold. The hoses
follow the pump.
-
Remove the NIBP pump.
-
Disconnect the two (T6) screws holding the manifold to
the front cover unit.
-
Lift the manifold carefully aside. Be careful not to
damage the NIBP flex board. Disconnect the NIBP flex
board connector from the STP input board.
NOTE: When reassembling, make sure that the NIBP flex
board connector is connected properly (all pins
connected) to the STP input board.