B30 Patient Monitor
6

Service Log

Error, event and alarm data is stored in the Service
Log.
The service log contains information about the
occurred monitor errors, events and alarms since
the last factory reset or service log reset. The service
log is saved in the EEPROM memory of the main CPU
board.
Error Log is for selecting the error history view onto
the right side of the menu. Error Log shows also
some monitoring events like warm and cold starts.
Alarm Log is for selecting the alarm history view
onto the right side of the menu.
Scroll Log is for scrolling the error/alarm
information on the right side of the menu.
Record Log is for recording the service log
information onto the recorder.
Reset Log is for clearing up the content of the
selected service log. This function should be run
after a performed maintenance.