To remove the NIBP pump and the
manifold unit
-
Disconnect the module bus connector, pump
connector and NIBP flex connector.
-
Disconnect the hoses (2 pcs) coming from the
manifold.
-
Lift up the NIBP board.
NOTE: Note the positions of the hoses; mark them if
necessary to ensure they are replaced correctly.
-
Disconnect the air intake hose from the NIBP manifold.
-
Remove the NIBP filter.
E-PSM(P)W module disassembly
Document no. 2044677-001
6-25