B30 Patient Monitor
4
E-PSM(P)W module disassembly

4.1 Before disassembly

NOTE: Wear a grounded, antistatic wristband when handling PC boards. Electrostatic
discharge may damage components on the board.
NOTE: Handle all PC boards by their edges.

4.2 Tools needed

CAUTION
When reassembling the module, make sure that all cables are reconnected
properly.

4.3 To disassemble the module

To open the E-PSM(P)W module
To remove the NIBP board
6-24
.
Document no
2044677-001
-
torx screwdrivers; T6, T8
-
flat blade screwdriver
-
pincers
-
antistatic wristband
1. Remove the four screws (T8) holding the module cover
to the frame from the bottom of the module.
2. Remove the two screws (T6) holding the lock unit to
the frame. While pulling the tab push the lockers with
a screwdriver to remove the lock unit.
3. Using screwdrivers to push the rubber snap, and open
the cover at the same time.
4. Hold the cover from the back corners, lift it about 45°
to unlock the snaps from the front unit and pull the
cover out backwards.
NOTE: Be careful not to damage the seal. When
reassembling the seal may stick to the cover.