Fcrec Module Disassembly And Reassembly; Instructions After Replacing Minic Unit; Before Disassembly - GE B30 Technical Reference Manual

Patient monitor
Hide thumbs Also See for B30:
Table of Contents

Advertisement

B30 Patient Monitor
5
N-FCREC Module disassembly and
reassembly
CAUTION
When reassembling the module, make sure that the tubes and cables are not
pinched between the boards and the cover.

5.0.1 Instructions after replacing MiniC unit

After replacing the MiniC unit:
perform the sampling system leak test
perform the occlusion test
perform the gas calibration

5.0.2 Before disassembly

NOTE: Wear a grounded, antistatic wristband when handling PC boards. Electrostatic
discharge may damage components on the board.
Disconnect the Extension Module from the monitor.
Tools needed
Disassemble the Extension Module
8-18
.
Document no
2044677-001
-
screwdrivers, TORX; T8, T10, T15
-
crosshead screwdriver; M15
-
flat blade screwdriver
-
pinchers
-
antistatic wristband
1. Remove the Mini D-fend.
2. Open the recorder unit paper cover and remove
the paper roll, if installed.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents