2
E-PSM(P)W Specifications

2.1 General specifications

Module size
W x D x H
Module weight
Operation temperature

2.2 Typical performance

2.2.1 NIBP

NOTE: Non-invasive blood pressure measurement is intended for patients weighing over 5 kg
(11 lb.)
Oscillometric measurement principle.
Measurement range
Pulse rate range accepted
Typical measuring time
Inflation limit
Venous stasis
Venous stasis pressure may be lower than the values above if the patient has low blood
pressure. The venous stasis pressure adapts to the measured mean pressure being the same
as mean pressure but always at least the following:

2.2.2 ECG

Lead type
Sweep speeds
Defibrillation protection
Recovery time
Input impedance
Allowable offset
Gain range
Display filter
Diagnostic
Monitoring
55 x 135 (175 w/ tab) x 145 mm
2.2 x 5.3 (6.9 w/ tab) x 5.7 in
0.6 kg /1.4 lb.
5 to 40°C / 41 to 104°F
adult
25 to 260 mmHg
child
25 to 190 mmHg
infant
15 to 140 mmHg
30 to 250 bpm
adult
< 30 s
infant
< 25 s
adult
280 ±10 mmHg
child
200 ±10 mmHg
infant
150 ±5 mmHg
adult
80 ±5 mmHg / 2 min.
child
60 ±5 mmHg / 2 min.
infant
40 ±5 mmHg / 1 min.
Infant 20 ± 5 mmHg
Child 30 ± 5 mmHg
Adult 40 ± 5 mmHg
3-Lead (I, II, III)
5-Lead (I, II, III, aVR, aVL, aVF, V)
12.5, 25, 50 mm/sec.
5000 V, 360 J
5 s
>2.5 MΩ (10 Hz)
±1VDC
0.2 to 5.0 cm/mV
0.05 to 150 Hz
0.5 to 30 Hz (-3 dB, with 50 Hz reject filter)
0.5 to 40 Hz (-3 dB, with 60 Hz reject filter)
E-PSM(P)W Specifications
7-3
Document no. 2044677-001