16. Use a flat blade screwdriver to unlock the module
bus connector insulator cover. Put the screwdriver
in the hole and move the blade backwards (away
from the module bus flex cable) until the insulator
cover unlocks.
17. By assisting with a screwdriver pull the module bus
connector through the hole in the module cover.
18. Flip the module cover over and remove the two (T6)
screws holding the lock unit to the frame. While
pulling the tab push the lockers with a screwdriver
to remove the lock unit.