3.1.2 Troubleshooting for ECG and STP parameters

P ossibly fault in m odule w ith
param eters.
C heck m odule configuration in S ervice
m enu
N O
Is it correct?
Y E S
Insert m odule and
turn pow er on
D oes fault
N O
still
appear?
Y E S
E nter the
S ervice m enu
N IB P
N O
m odule ID on
screen?
Y E S
S T P
N O
m odule ID
on screen?
Y E S
E C G m odule ID
on screen?
Y E S
C h e c k fro n t p a n e l
k e y fu n c tio n s in
S e rvic e m e n u
Y E S
D o
th e y w o rk ?
N O
C h e ck fro n t p a n e l u n it,
E C G b o a rd , a n d S T P b o a rd
to fin d c u lp rit. R e p lac e if
S elect the correct
configuration
F ault not in
m odule
N O
D oes another
Y E S
R em ove
m odule w ork in
N IB P board
sam e slot?
N O
H a s E C G
N O
H a s S T P
m ea s u re m e n t
b o a rd fa ile d ?
fa ile d ?
Y E S
Y E S
R e p la c e
R e p la c e
E C G b o a rd
S T P b o a rd
D oes another
Y E S
R em ove
m odule w ork in
E C G board
sam e slot?
H a s in v B P
N O
m e a s u r e m e n t
f aile d ?
Y E S
R e p la c e
S T P b o a rd
E-PSM(P)W Troubleshooting
R eplace S T P
board
N O
S T P m odule ID
on screen?
Y E S
R eplace E C G
board
H a s T e m p
N O
N O
m e a s u re m e nt
m e a s u re m e n t
fa ile d ?
Y E S
R e p la c e
S T P b o a rd
Document no. 2044677-001
H a s S p O 2
fa ile d ?
Y E S
R e p la c e
S T P b o a rd
4-9