Psm(P)W Troubleshooting; Troubleshooting Flowcharts; Troubleshooting For Nibp Parameter - GE B30 Technical Reference Manual

Patient monitor
Hide thumbs Also See for B30:
Table of Contents

Advertisement

B30 Patient Monitor
3
E-PSM(P)W Troubleshooting

3.1 Troubleshooting flowcharts

3.1.1 Troubleshooting for NIBP parameter

START
Insert new module and turn
power on.
Does fault still appear?
YES
Module ID on screen?
YES
Error message on
screen?
YES
See error code explanation in
service manual and fix it.
Do they work?
NO
Check tubes and
connectors.
Find leak and fix it.
4-8
.
Document no
2044677-001
NO
Does another
NO
module work in
same slot?
NO
Fault not in original
module
NO
Key check in
Service Menu OK?
YES
Pump check in
Service Menu OK?
YES
Leak test in Service
Menu OK?
NO
YES
NO
Check keyboard
connector. If
OK, change
front panel.
Check pump
NO
connector.
If OK, change
pump.
YES
Start NIBP
without hose
Cuff ID OK but measurement
Replace
the module.
NO
OK?
Check and
replace
NIBP board.
NO
OK?
Check and replace
- STP Input Board
- NIBP flex board
- connectors
Cuff loose
NO
message appears
on screen?
YES
does not work