6.8.4. Mitigation of risks due to disconnection from the tablet PC  
When  using  the  GEMS‐H,  status  monitoring  and  operation  are  done  through  a  tablet  PC  via  Bluetooth  wireless 
communication.  If  the  Bluetooth  communication  is  interrupted,  the  monitoring  and  control  of  the  GEMS‐H  are 
designed to cease, but the device will continue operation.  
 
Status of the device and recommended user action:  
In the case that communication ceases, the tablet PC alerts the physical therapist by displaying the "DISCONNECTED" 
 
warning message on the application. 
For safety, when such an event occurs, the GEMS‐H continues at the current assistance setting. The physical therapist 
should stop the current treatment and reconnect the tablet PC. When the tablet PC is reconnected to the GEMS‐H, 
the application will display the message "CONNECTED", and the tablet PC resumes operation.  
 
Error message examples 
 
 
  
[Warning message] 
[Screen of connected situation] 
 
 
  
SAIT‐IFU‐GEMS‐H Rev.1.1 (2019‐11) 