5.2.3. Size markings
The size of the waist part is indicated on the front clasp (S, M, and L).
5.3. Adjust the waist part width
Property damage when the screws are over tightened!
The product may be damaged if the screws are over tightened or continuously tightened after clicking sounds
are emitted from the torque wrench provided.
‐
Do not tighten further after clicking sounds emit from the torque wrench provided.
Property damage when the screws are not fully tightened!
If the screws are not fully tightened, the screws will be too loose during walking and the waist part may be
damaged.
‐
Before putting the device on the patient, check that the screws are tightened securely enough that noise
is not made, as the screws loosen during walking.
① Ensure the Velcro on the waist belt is unfastened.
② Adjust the device width by loosening the two screws, one on each side of the control pack.
SAIT‐IFU‐GEMS‐H Rev.1.1 (2019‐11)