5.7. Check prior to use/operation of the device  
5.7.1. Check for before use  
① Check that the screws for adjusting the hip width of the waist part are not loose. 
② Check that the thigh support frame fastening hook and strap Velcro are securely fastened. 
③ Use the device after it is fully charged. Monitor the battery level and if the battery level is below 20 %, charge 
the device before use. 
 
 
5.7.2. Check for any discomfort while walking  
 
 
① Before turning the device on, make the wearer walk around to check for any discomfort. If there is discomfort, 
reposition the device or try donning the device again. You may need to readjust the width of the waist part or 
the waist belt tightness.  
② Confirm  with  the  patient  that  there  is  no  pinching  of  the  skin.  If  the  patient  is  comfortable  then  the 
measurement is acceptable.  
③ Check  if  the  greater  trochanter  area  is  making  contact  with  the  hard  surface  of  the  device.  This  could  cause 
discomfort after walking for some time. If this is the case, the device width may need to be widened.  
④ Avoid  wearing  the  device  with  short  pants  where  the  skin  is  exposed  to  the  thigh  support  strap  area.  When 
walking a long time, this could cause skin chafing. 
 
 
 