6.7.2. Assist control area 
6.7.2.1. Set the stand pose 
 
Make sure that hip angle is as close to neutral as possible with the patient standing naturally. If the patient is 
walking with their back bent, their neutral posture will also be standing with back bent.  
Incorrect setting of the stand pose may cause uncomfortable assistance or degrade clinical effectiveness.  
 
 
Make sure that both the left and right joint angle options are selected within the monitoring option dialog.  
 
① Reset the hip angle sensor while the patient is standing in a neutral posture with hips straight if possible, or with 
comfortable posture otherwise. This will be used as the reference position for calculating flexion and extension.  
② Press the "Set Stand Pose" button to reset the hip angle to zero and to calibrate the reference posture.  
③ If the stand pose is not set, the "START" button will not be activated.  
④ Make  sure  that  both  left  the  right  joint  angles  are  selected  in  the  monitoring  option  dialog  to  see  both  joint 
angles in the graph.  
 
SAIT‐IFU‐GEMS‐H Rev.1.1 (2019‐11) 
 
 
 
 
Page 45 of 79