6.8.2. Mitigation of risks due to excessive power output  
In order to avoid an excessive torque being applied to the patient, the device is designed to not to transmit a torque 
of  more  than  12  Nm  ±15  %  to  the  patient.  When  the  assistive  force  reaches  the  predetermined  maximum  force, 
then the device will cease applying the assistive force. 
 
Status of the device and recommended user action:  
If assistive force is ceased due to an over‐torque situation, the associated tablet PC will display an error message of 
"EXCEEDED TORQUE LIMIT."  
When this message is displayed and the device is stopped, the physical therapist may try to start the device again by 
pressing the "START" on the tablet PC. If the problem persists, reboot the device and the tablet PC. If there is still the 
problem with the device, turn off the device immediately and contact the manufacturer. 
 
Error message examples 
 
 
 
SAIT‐IFU‐GEMS‐H Rev.1.1 (2019‐11) 
Page 55 of 79