When Pcb-50Led (Bg7-2767) Is Replaced; When Pcb-50Ad (Bg7-2770) Is Replaced; When Pcb-50Xray (Bg7-2770) Is Replaced; When 50 Power Unit 7M (Bg7-2857) / 3M (Bg7-2858) Is Replaced - Canon CXDI-50G Service Manual

Cxdi-1 system
Hide thumbs Also See for CXDI-50G:
Table of Contents

Advertisement

4. Repair Guide
4) Install the sensor serial number.
Refer to "(4) Checking the Sensor Serial Number" in "5.2.5 Settings" in "2.5 Installation" of "2.
Installation Manual".
5) Install the sensor data file.
Refer to "DP File Write Tool Software" in the Tool Software Operation Manual for CXDI-50G.
To perform this operation, "Sensor Data File FD" that comes with the sensor unit is required.
6) Check the image quality.
Refer to "5.2.6 Image Quality" in "5.2.5 Settings" in "2.5 Installation" of "2. Installation
Manual".

(3) When PCB-50LED (BG7-2767) is replaced

1) Check that the LED lights.
2) Check the image quality.
Refer to "5.2.6 Image Quality" in "5.2.5 Settings" in "2.5 Installation" of "2. Installation
Manual".

(4) When PCB-50AD (BG7-2770) is replaced

1) Check the image quality.
Refer to "5.2.6 Image Quality" in "5.2.5 Settings" in "2.5 Installation" of "2. Installation
Manual".
3.2
Power box

(1) When PCB-50XRAY (BG7-2770) is replaced

1) Set the jumper pins on the PCB.
Refer to "(2) Power Box" in "2 PCB settings" in "4. Repair Guide".
2) Check that the system starts up normally and that operations such as exposure, images
capture and data transfer are performed normally.

(2) When 50 Power Unit 7m (BG7-2857) / 3m (BG7-2858) is replaced

1) Check that the system starts up normally and that operations such as exposure, images
capture and data transfer are performed normally.