Error Code
Reply
Seriousness
Keys
-311
OK
Error
-402
OK
Error
-403
OK
Error
Error Code List for Version 6.00
[Operation after Reply Key is touched]
#1:Error
#2:REX Value Warning
#3:REX is 9999. It is under the maximum limitation value
999.
Please reconfirm the generator setting or exposure
condition.
If the condition is correct, adjust the image processing.
[Cause]
This error occurs when the REX value acquired by image
processing after the imaging is more than the limitation value.
[Operation after Reply Key is touched]
Nothing
[Note]
9999: REX value obtained by image processing
999: upper limit value
#1:Error
#2:Incomplete End Error
#3:The last image in the last study was not saved completely.
[Cause]
Failed to write in the last image of the previous imaging.
[Operation after Reply Key is touched]
If there remain the previous images, they will be recovered.
#1:Error
#2:Incomplete End Error
#3:Do you want store captured images into internal storage in
order to recover them later?
(Name of the patient)
(Number of images)
[Cause]
The last image(s) was not queued due to incomplete end in
the previous imaging.
[Operation after Reply Key is touched]
When Ok is chosen, the recovery process will start. When
Cancel is chosen, the last images will be deleted. In this case,
normal exposure for imaging will start without recovering the
last images. When CANCEL is chosen, a confirmation
message will appear before deleting the image.
- 11 -