be used with the generator, the probability of the warning message is very low.)

3. Adding values to the imaging conditions table

It is possible to add values up to 3000 msec, 4800 mAs to the imaging conditions table.
However, at the time of the actual imaging, if the exposure is performed with a CXDI-11 -
CXDI-31 sensor, it is not possible to set values higher than 1000 msec, 1600 mAs.

4. Imaging conditions table

It is possible to add values up to 3000 msec and 4800 mAs to the system settings -
imaging conditions table on the settings screen. However, if 1001 - 3000msec, 1601 -
4800mAs values are added, a warning message will be displayed only once, saying the
values cannot be used with the CXDI-11 - CXDI-31 sensors. [F
If at the time of a regular exposure or a calibration exposure, the imaging body part of a
CXDI-11 - CXDI-31 sensor was selected, values higher than 1000 msec, 1600 mAs
cannot be selected from the imaging conditions table.
3. Function
Figure 3-1
Figure 3-2
6
igure 3-2
]
Press [OK]