5 INSTALLATION

EC Directive 89/336/EEC
In order to meet the requirements of EC Directive
89/336/EEC for EMC regulations, this product must not
be used in a non-industrial environment.
End of Life Disposal
• The instrument contains a small lithium battery
which should be removed and disposed of
responsibly in accordance with local environmental
regulations.
• The remainder of the instrument does not contain
any substance that will cause undue harm to the
environment and may therefore be considered as
normal waste and disposed of accordingly.
Cleaning
Clean only the front panel, using warm water and a mild
detergent.
5.1 Siting – Figs. 5.1 and 5.2
Keep distance
to a minimum
A – Close to Sensors
B – At Eye-level Location
C – Avoid Vibration
Fig. 5.1 General Requirements
100
Sensors
0C
(32F)
Min.
A – Within Temperature Limits
0 to 95% RH
B – Within Humidity Limits
IP66 / NEMA 4X front
IP40 rear (with extension)
IP20 rear (without extension)
C – Environmental Limits
+
Warning. Select a location away from
strong electrical and magnetic fields. If this is not
possible, particularly in applications where
mobile communications equipment is expected
to be used, screened cables within earthed
metal conduit must be used.
D – Use Screened Cable
Fig. 5.2 Environmental Limits