Honeywell ST 3000 User Manual page 187

Smart transmitter release 300 and smart field communicator model sts103
Hide thumbs Also See for ST 3000:
Table of Contents

Advertisement

9.3
Inspecting and Cleaning Barrier Diaphragms,
Procedure, continued
6/05
ST 3000 Release 300 and SFC Model STS103 User's Manual
Table 53
Inspecting and Cleaning Barrier Diaphragms, continued
Step
6
Coat threads on process head bolts with anti-seize compound such
as "Neverseize" or equivalent.
7
Replace process head or heads and bolts. Finger tighten nuts.
8
Use a torque wrench to gradually tighten nuts to torque rating shown
in Table 48 in sequence shown in following illustration. Tighten head
bolts in stages of 1/3 full torque, 2/3 full torque, and then full torque.
Always tighten head bolts in
sequence shown and in these
stages:
1. 1/3 full torque
2. 2/3 full torque
3. Full torque
9
Return transmitter to service.
CAUTION Do not exceed the overload rating when placing the
transmitter back into service or during cleaning operations. See
Overpressure ratings in Section 3 of this manual.
Action
1
4
Continued on next page
Continued
3
2
173