Event
Description
ID
232
Internal voltage error
233
Tape advance failure
234
Internal Comm failure
Abnormal Az current consumption is
235
detected
Internal fault at control module
236
Internal fault at PDU module
237
238
Analyzer SW corrupted
239
Internal SW fault at analyzer
240
Pump failure
241
IO panel not responding to Analyzer
242
Internal fault at IO panel module
243
FAN fail
Vertex
TM
C System
Possible Cause
Internal voltage hardware issue
Encoder failure
Cable issue
Stepper Motor Failure
Cable issue
Hardware failure
Hardware failure
Hardware failure
Hardware failure
Software installation failure
Internal SW operation failed
Line voltage issue
Pump failure
Circuit breaker tripped
Leakage in system
Cable issue
IO panel SW operation failed
IO panel SW operation failed
Fan blade gets stuck
Fan short circuit
Contact Honeywell Analytics Service
Replace Encoder
Check Encoder cabling
Contact Honeywell Analytics Service
Replace Stepper Motor
Check Stepper Motor cabling
Contact Honeywell Analytics Service
Contact Honeywell Analytics Service
Contact Honeywell Analytics Service
Contact Honeywell Analytics Service
Contact Honeywell Analytics Service
Re-install analyzer software
Replace Analyzer, Contact Honeywell service
Reboot the Analyzer. Contact Honeywell service.
Verify main line voltage
Replace pump, Contact Honeywell service
Contact Honeywell service
Check tubing and fitting in system, Contact Honeywell service
Check IO panel cabling
Reboot the IO panel module. Contact Honeywell service.
Reboot the IO panel module. Contact Honeywell service.
Check fan blade, Contact Honeywell Analytics Service
Replace fan, Contact Honeywell Analytics Service