Atex Certified Advanced Hand Terminal - GE Sensing Druck ADTS 2 Series Calibration Manual

Air data test set
Hide thumbs Also See for Sensing Druck ADTS 2 Series:
Table of Contents

Advertisement

Druck ADTS 2XX Calibration Manual
CONDITIONS OF USE
The ATEX certified Advanced Hand Terminal can be used in zone
accordance with the ATEX certification document and schedule.
Marking detail:
Refer to Advanced Hand Terminal User Manual K0418 and the label on the Advanced Hand
Terminal.
SPECIAL CONDITION OF USE
The power supplies must be isolated when connecting the advanced hand terminal in the
hazardous area.
The advanced hand terminal must not be disconnected when energized in the hazardous
area.
The advanced hand terminal is a non-serviceable component. If the advanced hand terminal
becomes unserviceable it can only be replaced by another ATEX compliant hand terminal.
Note:
The advanced hand terminal must only be used with the cable assembly supplied and marked "DO NOT
SEPARATE WHILST ENERGISED IN HAZARDOUS AREA"
*
Zone 2 hazardous area definition, see User Manual K0417 or User Manual K0451.
K0425 Issue No. 2
ATEX C
A
ERTIFIED
DVANCED
H
T
AND
ERMINAL
2*
hazardous areas in
iv