How to Replace the Battery
Revision A
Maintenance: Lithium-Ion Battery
1. Open the battery door by gently pulling on the battery door pull tab.
2. Pull on the battery pull strap to remove the battery from the monitor.
3. Insert the new battery.
a. Insert the connector end of the battery into the battery slot.
Verify the connector ends are facing the back of the monitor.
b. Push the battery firmly into the slot.
4. Gently push the battery door closed until it "snaps" into position.
127(
The battery door will not close if the battery is not fully inserted
into the battery slot.
5. Verify that the monitor operates correctly.
a. Confirm that the battery icon is displayed in the lower right
corner of the monitor.
b. Verify the green or amber battery CHARGING STATUS LED is
illuminated.