Table 4-3 Indicator Unit Group Function Prompts - Honeywell 7866 Operation And Maintenance Manual

Triple range digital h2 and co2 analyzer/indicator and gas sampling panel
Table of Contents

Advertisement

Function Prompt
(Lower Display)
SECURITY
Security Code
LOCKOUT
Configuration Lockout
Analyzer Range Type
TYPE
BIAS R1
Bias of Input Value for Range 1
BIAS R2
Bias of Input Value for Range 2
BIAS R3
Bias of Input Value for Range 3
PV INPUT
Selection of Input actuation
consistent with analyzer CUR
OUT selection
RNG3SPAN
Selection of Input Range 3 to be
consistent with similar function
in the Analyzer.
For Range 3, it applies 1-5V or
0-5V over the range of 0-100%,
75-100%, or 100 - 75%.
Input Frequency Selection
PWR FREQ
DECIMAL
Decimal Point Location
9/06
7866 Triple Range Digital H

Table 4-3 Indicator Unit Group Function Prompts

Function Name
and CO
Analyzer/Indicator - Operation and Maintenance
2
2
Selections or Range of Setting
(Upper Display)
0000 to 9999
NONE – Permits changes to all functions.
CALIB – Calibration group functions are
hidden from view.
UNIT – Parameters in the Unit Set Up group
cannot be changed, but the Alarms and
Communications group parameters are
changeable. (Calibration parameters cannot
be seen.)
VIEW – All parameters are read only.
(Calibration parameters cannot be seen.)
TRIPLE – Read only
INDICTR – Read only
+10.00 to –10.00
+10.00 to –10.00
+10.00 to –10.00
1-5V (4-20mA)
0-5V (0-20mA)
0 - 100%
75 - 100%
100 - 75%
60 HZ
50 HZ
XXX.X – One decimal place
XX.XX – Two decimal places
Set Up Mode
Factory
Setting
0
NONE
TRIPLE
0.00
0.00
0.00
1-5V
0-100
60 HZ
XXX.X
35