7. Technical Specification

7.1. LAB-X5000 Series

Device type code
Rated voltage
Rated current
Rated frequency
Fuse
Average power consumption
Operating temperature range
Storage temperature range
Recommended storage temperature
Operating humidity range
Maximum operating altitude
Embedded storage
Available storage space
Maximum measurement count
Dimensions (W x L x H)
Weight
Pollution Degree
Ingress Protection degree
Overvoltage category

7.2. X-Ray Tube

Target material
Window thickness
Anode voltage range
Maximum anode current
Maximum anode power
Nominal filament voltage
Maximum filament current
Minimum counts
Ambient temperature range
Maximum package temperature
LXDS2860
100 V AC to 240 V AC
1.5A
47 Hz – 63 Hz
T 2A H 250 V
60W
+10 ˚C to +35 ˚C / +50 F to + 95 F
-10 ˚C to +50 ˚C / +14 F to + 122 F
Room temperature
20 % RH to 80 % RH
2000 m / 6500 ft
NAND 4 GB Flash Memory
2.3 GB
100 000
440 mm x 520 mm x 155 mm
20 kg
PD2
Not classified
II
Palladium (Pd) or Titanium (Ti)
0.005 in
4 kV to 30 kV
0.75 mA
3.0 W
2.5 V
1.3 A
24000
1/s
0˚C to +40 ˚C / +32 F to + 104 F
55 ˚C / +131 F