LAB-X5000 Series User Guide
5.4.3.2.
Standards can be certified reference materials, samples that have been
well-analyzed using another analytical technique, or synthetics.
5.4.4.
Using the chosen standards, a calibration can be derived by specifying the instrument
measurement conditions, then measuring the standards and performing regression
calculations, which may involve instrument or matrix corrections.
5.4.5.
Once a calibration has been established, it can be saved for future recall and analysis,
or can be exported to another LAB-X instrument.
5.4.6.
To create a new calibration: from the pull-down menu, select Method, then select New
Calibration.
5.4.6.1.
Type in a name for the new calibration, then tap
to the Calibration menu.