5.5.2.1.
Intensities: A display of the original (raw) data and the corrected intensity used to
perform the regression.
(Blank, Overlap) are applied.
5.6. Import Calibration
5.6.1.
Measuring a set of calibration standards using a Hitachi pre-programmed calibration
5.6.1.1.
Pre-programmed calibrations cannot be edited or deleted.
5.6.1.2.
In order to use a pre-programmed calibration as the basis for a new one, it is
necessary to copy the pre-programmed calibration and save it as a new calibration name,
then measure an appropriate set of standards and SUSs.
5.6.2.
Importing a calibration from another LAB-X instrument.
5.7. Test Measurement
5.7.1.
Test Measurement allows for a full spectrum scan of a sample, after which the resulting
spectrum scan can be investigated (i.e. element peaks identified, view is zoomed in or
out, saved to USB, or a second sample can be measured and the two spectra
superimposed.)
5.7.2.
Select Method from the pull-down menu, then tap Test Measurement.
5.7.3.
Sample conditions are configured prior to analysis, such as excitation conditions, filter,
and analysis time.
Corrected intensities are used when instrument corrections
Hitachi High-Tech Analytical Science
▶
35