INTELLIDOX DOCKING MODULE
Bump Test Guidelines
1.
Perform compliance tests only in a normal environment that is 20.9% v/v O
of hazardous gas. Do not operate the module in a hazardous area. Failure to adhere
to this guideline can result in possible personal injury and/or property damage.
2.
The use of gas cylinders other than those specified by Honeywell may result in
unsafe bump tests or possible non-recoverable failure of the equipment, and will
invalidate the warranty.
3.
Do not use gas cylinders past their expiration date.
4.
A demand flow regulator must be used with all gas cylinder connections. The module
inlet should not be pressurized.
5.
Do not bump test the detector immediately after charging is complete.
6.
Bump test the detector if ambient gas readings vary during startup.
7.
Allow the sensor to stabilize before starting the bump test.
a)
b)
8.
When bump testing the same detector multiple times, wait 10 minutes between bump
tests to allow the sensors to stabilize.
HONEYWELL
Used sensor: wait 60 seconds
New sensor: wait 5 minutes
USER MANUAL || DETECTOR OPERATIONS
and free
2
PAGE 49 OF 119