INTELLIDOX DOCKING MODULE
Transferring Detector Firmware via Safety Suite Device COnfigurator
When a module or group of modules is connected to a network, you may transfer the
detector firmware update file to multiple selected modules via Safety Suite Device
Configurator software. For more information, refer to the Safety Suite Device Configurator
operator manual.
1.
Download and save the firmware update file to a PC or network drive. Do not
rename the file.
2.
Start Safety Suite Device Configurator software, and then log in as an
administrator. Select Configure Devices via IntelliDoX on the Devices menu.
The IntelliDoX Configuration Selection dialog box is displayed.
3.
Select a configuration file and click OK. The IntelliDoX Device Configuration
dialog box is displayed.
4.
Click Bootloader. The IntelliDoX Selection dialog box is displayed.
5.
Click Choose File. The Choose Firmware File To Upload dialog box is
displayed. Select the downloaded file, and then click Open. The IntelliDoX
Selection dialog box is displayed.
Select one or more IntelliDoX modules, and then click OK. A progress screen is
6.
displayed while the firmware update file is automatically transferred to the selected
modules.
7.
When the transfer is complete, click OK to close the dialog box.
Updating the Detector Firmware via Safety Suite Device Configurator
Modules must be configured via Safety Suite Device Configurator software to enable
firmware updates on insertion for compatible detectors. For more information, see
Configure Module and Detector Settings.
1.
Insert a compatible detector into the module. Detector identification is displayed
on the LCD.
2.
If the module is configured to perform a firmware update and other procedures on
insertion and a firmware update is available, then the detector event logs are
automatically transferred to the module and the firmware update begins.
3.
The LCD changes to yellow. A progress bar and the Detector firmware upgrade
message are displayed.
4.
After the upgrade is completed, the detector reboots. Alternating Detector
rebooting and Do not open lid messages are displayed.
When the reboot is successful, the LCD changes to green and the Firmware
5.
upgrade complete message is displayed. Other procedures specified to occur
automatically on insertion are performed.
6.
After all operations are completed, verify the firmware update.
HONEYWELL
a)
Open the lid and remove the detector.
b)
Verify the firmware version. For more information, refer to the
appropriate detector operator manual.
USER MANUAL || UPDATE FIRMWARE