INTELLIDOX DOCKING MODULE
Removing an Individual Module from Parallel DIN Rails
1.
Disconnect the power cord, network cable, tubing and inlet filter assemblies.
2.
Remove the end plate.
3.
Find the DIN mount lock on the lower right side of the module.
4.
Hold the DIN mount lock down, and then tilt the module away from the lower DIN rail until
the lower DIN mount is released. Release the DIN mount lock and remove the module
from the upper DIN rail.
Removing Connected Modules from Parallel DIN Rails
1.
Disconnect the power cord, network cable, tubing and inlet filter assemblies.
2.
Remove the end plate.
3.
Lift the latch arm between connected modules at the end of the group.
4.
Find the DIN mount lock on the lower right side of the last module. Hold the DIN mount
lock down, and then slide the module along the DIN rails until it is separated from the
group.
5.
Hold the DIN mount lock down, and then tilt the module away from the lower DIN rail until
the lower DIN mount is released. Release the DIN mount lock and remove the module
from the upper DIN rail.
6.
Repeat steps 3, 4 and 5 until only one module remains.
7.
Repeat steps 3 and 4 to remove the final module.