4.
From the ACTIONS column, expand ACTIONS and select Update Passcode.
5.
Enter a 6-digit network Passcode, and then click Save. The Network passcode status is
displayed. The IntelliDoX is added, and the Safety Suite Device Configurator can access
data.
Connect a Calibration Gas Cylinder
The IntelliDoX Enabler Kit includes tubing and quick connect fittings that are appropriate for
use with calibration gas cylinders that are approved for use with this product. For best results,
Honeywell recommends that the tubing is between 39 inches (1 meter) and 33 feet (10 meters)
in length for calibration gas cylinders.
Before You Begin
1.
Read and understand the warnings and cautions in this section.
2.
Verify that you are in a normal atmosphere (20.9% v/v O
3.
Verify that the end plate is attached and the latch arm is lowered and locked.
4.
Verify that the exhaust tubing is connected, and that it is free of defects and obstructions.
The maximum recommended calibration tubing length is 10 meters (33 feet). Use of
tubing that is longer than the recommended length may result in inaccurate or failed
compliance tests or calibrations.
A demand flow regulator must be used with all calibration gas cylinder connections. The
module inlet should not be pressurized.
The use of calibration gas cylinders other than those specified by Honeywell may result in
unsafe calibration or possible non-recoverable failure of the equipment, and will invalidate
the warranty.
Do not use calibration gas cylinders past their expiration date.
Ensure that all calibration gas cylinders are in good condition.
Ensure that all calibration gas cylinders contain enough gas.
This equipment uses potentially harmful gas for calibrations. The equipment must be
attached to a venting system or be used in a well-ventilated area.