INTELLIDOX DOCKING MODULE
SAFETY SUITE DEVICE CONFIGURATOR AND SAFETY SUITE SOFTWARE
e)
f)
Verify that the configuration settings are updated.
g)
Repeat steps 3 through 6 until configuration settings are verified for all modules.
Transferring the Configuration File to a Module via USB Flash Drive
For connected modules, transfer the configuration file to each appropriate
module in the group.
Note: during update through USB two messages can be shown "Communication
Failure" and/or "Configuration Corrupt". These messages may appear only while the
IntelliDoX is in a gang arrangement. If one IntelliDoX in this arrangement displays the
warning message "CONFIGURATION CORRUPT SETTINGS LOADED FROM FILE" it
is likely that another IntelliDoX will show the "MULTI-DOCK COMMUNICATION
FAILURE" warning message since the devices have different configuration
parameters. To make the message disappear press the check mark button to remove
the message from the screen.
1.
Insert a USB flash drive into a USB port on your PC. Use Windows Explorer or My
Computer to open the USB flash drive, and then create a new folder named System.
2.
Create the configuration file. See Error! Reference source not found. Safety Suite
Device Configurator or Safety Suite.
3.
Click Save to File. The Save Settings to File dialog box is displayed.
4.
Save the configuration file to the System folder on the USB flash drive. Do not
change the file name or extension. Eject the USB flash drive from the PC.
Insert the USB flash drive into the USB port on the module. The USB File Copy
5.
Utility menu is displayed. Press
continue.
The LCD changes to yellow. A progress bar and alternating Do not remove stick
6.
and USB transfer in progress messages are displayed while the file is transferred.
7.
After the transfer is successfully completed, the LCD changes to blue and the
USB File Copy Utility menu is displayed. Press
press
8.
The LCD changes to green. Remove the USB flash drive from the port when
prompted.
Transferring Configuration Settings to a Detector
1.
Insert a compatible detector into the module. Detector identification is displayed on
the LCD.
2.
If the detector is configured to update settings on insertion, then the detector settings
are updated automatically after the detector is recognized.
3.
After the configuration update is completed, other procedures specified to occur on
insert are performed.
4.
After all procedures are performed, remove the detector from the module and verify
that the configuration settings are updated. For more information, refer to the
appropriate detector operator manual.
HONEYWELL
After the settings are retrieved, the IntelliDoX Device Configuration dialog
box is displayed. The serial number of the selected module is appended to
the dialog box title bar.
or
to continue.
USER MANUAL || CONFIGURE SETTINGS VIA
to select Copy to dock and then press
or
to select Exit and then
to
PAGE 69 OF 119

Advertisement

Table of Contents