INTELLIDOX DOCKING MODULE
SAFETY SUITE DEVICE CONFIGURATOR AND SAFETY SUITE SOFTWARE
Setting the Module Time
1.
Start Safety Suite Device Configurator software, and then log in as an administrator.
2.
Select Configure Devices via IntelliDoX on the Devices menu. The IntelliDoX
Configuration Selection dialog box is displayed.
3.
Select a configuration file and click OK. The IntelliDoX Device Configuration dialog
box is displayed.
4.
Select the tab for a compatible detector, and then select the IntelliDoX
Configuration tab.
5.
Click on the Time zone selector. Select your time zone from drop down menu.
6.
Verify that the Current date and time displayed under Time zone settings is
correct. The date is synchronized with the PC on which Safety Suite Device
Configurator software is installed. If the date is incorrect, change the date on the PC.
If the time is incorrect, click on the Time zone selector to select a different time zone.
7.
When the correct date and time are displayed, click Set IntelliDoX Time. The
IntelliDoX Time Confirmation dialog box is displayed.
8.
Click Save to IntelliDoX. The IntelliDoX Selection dialog box is displayed. Select
one or more IntelliDoX modules, and then click OK.