Instructions For Reading Settings; Ce - Convert Error Code; Cid - Suppression Of Id Codes - AEG ARE i2 - HF Installation Manual

Compact reader
Table of Contents

Advertisement

5.3

Instructions for reading settings

5.3.1 CE – convert error code
With CE=1 the reader sends no error codes, except the no read error, during the md0 mode or the
commands Get Tag. The leds are not influenced by this command. This command has only effect in
the ISO 14443A standard.
This command is not available in profibus communication and can not be saved with the VSAVE
command.
Input format:
Output (example):
Parameter:
PARAMETER
0
1
5.3.2 CID – suppression of ID Codes
In the MD0 mode with CID=1 only the first of in succession identical transponder numbers is output
on the serial interface. The possibly following identical transponder numbers are suppressed, as long
as no new valid transponder number is received, processed and output. The get tag command is not
influenced by this command. NoReads do not influence the data filtering.
This command is not available in profibus communication and can not be saved with the VSAVE
command.
Input format:
Output (example):
Parameter:
PARAMETER
0
1
--------------------------------------------------------------------------------22/50--------------------------------------------------------------------------------
CE <SP> parameter <CR>
0 <CR>
FUNCTION
No suppression
Suppression of error codes
CID <SP> parameter <CR>
0 <CR>
FUNCTION
No suppression
Suppression of equal transponder numbers