Density Programming - GE Alpha RT 32551 Installation Instructions Manual

Table of Contents

Advertisement

32551-IMG rev 6
5.2.19

Density programming

1
Select the molybdenum filter.
2
Install the cassette holder.
3
Check that the semiAuto (A) mode is on. Make a test exposure with
the default kV using 40 mm acrylic. Measure the O.D. If the O.D. is not
correct, increase or decrease the density with Den+/- keys. Repeat
until the O.D. is correct.
4
Mark the density value that produces the right optical density in the
table 2.9 or 2.10 and store the values by programming them in the
program (Pr) mode.
If the film/screen combination and default kV is the same for all channels
a, b and c, the density correction is the same also.
5
Install the magnification tunnel.
6
Repeat the instructions 3. and 4.
7
Change the rhodium filter
8
Repeat the instructions 2. - 6.
NOTE!
In order to enter the Program (Pr) mode, press Density+ and Density-
buttons simultaneously.
GE Healthcare
5 Setups, alignments and adjustments
87

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents