3 Installation
7
Attach the shield decal enclosed.
Both side installation:
Use the upper hinge bracket with hinges on both sides. Follow the
instructions given for left or right side installation.
Fig 3.13.
Left side installation
3.9
M
AINS PLUG INSTALLATION
The unit is delivered without mains plug attached to allow fitting of locally
suitable type of mains connection.
Select suitable min 16A rated mains plug (that complies with local
regulations) with ground (PE) connection. A loose IEC type plug is included
in the unit delivery.
When connecting the wires, make sure that the yellow-green PE wire is
left min 10mm longer than the two other ones.