NVIRONMENTAL REQUIREMENTS
Operating Temperature
Humidity Range
Weight
Dimensions
NOTE!
Long term storing in dry cool place.
2.5
I
NSTALLATIONS REQUIREMENTS
CONTROL
In order to minimize interference risks, the following requirements shall
apply.
Cables shielding & grounding
All interconnect cables to peripheral devices must be shielded and
properly grounded. Use of cables not properly shielded and grounded
may result in the equipment causing radio frequency interference.
Stacked components & equipment
The Alpha RT should not be used adjacent to or stacked with other
equipment; if adjacent or stacked use is necessary, the Alpha RT should be
observed to verify normal operation in the configuration in which it will be
used.
Electrostatic discharges environment & recommendations
In order to reduce electrostatic discharges interferences, the floor
material shall be a dissipative material, versus the electrostatic charge
stacking. The relative humidity shall be at least 30 percent. The dissipative
material shall be connected to the system ground reference.
Pins of connectors should not be touched and connections should not be
made to these connectors unless ESD precautionary procedures are used.
2.6
T
OOLS NEEDED FOR INSTALLATION
– Electrical screwdriver (to open the crates)
– Screwdrivers small/medium
– Screwdrivers POZIDRIVE small/medium
– Allen key set (long type 1.5 - 8 mm)
– Wrench set 7 - 19 mm.
– Pliers
– Drilling machine with 10 mm bit for wall mounting
GE Healthcare
2 Preinstallation instructions
+10...+40°C (60° - 85°F)
15 - 85%, non-condensing
258 kg (568 lbs)
Height 224 cm (88.2")
Width 34.2 cm (13.5")
Length 101.5 cm (40")
&
ENVIRONMENT