Lifting The Column - Minimum 2 People; Installation Of The Column To The Base Plate - GE Alpha RT 32551 Installation Instructions Manual

Table of Contents

Advertisement

32551-IMG rev 6
3.2.3

Lifting the column - minimum 2 people

CAUTION!
Make sure that the counterweight is resting against the assembly pin and
the carriage is down so that the cables are taut.
1
Lift the column up smoothly so that the counterweight lies on the
assembly pin, and attach the wall bracket to the holes in the wall.
WARNING!
The column assembly is heavy. Get assistance and be careful when
erecting the column.
2
Attach the column to the wall with two 10 mm bolts.
3
Check with a water-level that the column is vertical in both
directions.
CAUTION!
Do not remove the rubber band, holding the cable coming out through the
axle because it helps in installing the C-arm by keeping the cables out of
the way.
3.3
I
NSTALLATION OF THE COLUMN TO THE BASE PLATE
1
Unpack the base plate and compare its contents to the packing list.
2
Place the base plate in its final position so that the screw holes face
to the rear of the unit.
3
Bolt both side pieces to the base plate making sure they are centered
over the bolt holes. Trace the inside edge of the side plates.
4
Remove enough screws to allow the side pieces to swing out of the
way.
5
Install the rubber bumper to the base plate.
CAUTION!
Ensure that the counterweight is resting against the assembly pin and the
carriage is down so that the cables are taut.
CAUTION!
Damage to the surface of the base plate can occur if the column is
attempted to "walk" or slide.
6
Move the column over to the base plate and stand it up on top of
rubber bumper and the traced outline.
GE Healthcare
3 Installation
21

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents