Setup Info Printout And Printer/Pc Connection #62 Inf - GE Alpha RT 32551 Installation Instructions Manual

Table of Contents

Advertisement

5 Setups, alignments and adjustments
The kV is selected on the basis that the target time for the compressed
breast thickness will be reached. The higher target time will result in lower
selected kV and vice versa.
To enter the service mode:
To enter the service mode press kV- and density- buttons simultaneously
until the Timer display shows Ser.
To enter the special program menu:
To enter the special program menu press density+ and density- buttons
simultaneously. The program number is displayed in the kV display and
the program name in the mAs display (Press DEN- first because otherwise
the system may toggle the focal spot with DEN+ key). The available
programs can be scrolled with the kV+ and kV- buttons. To enter the
desired program press the Autoexp program button. To leave the
program press Autoexp buttons once more. To leave the menu press
density+ and density- simultaneously.
Return to the menu from the special programs by pressing AUTOEXP
button.
NOTE!
See block diagram of software in appendix or service manual.
5.1.2

Setup Info Printout and printer/PC connection #62 InF

Connect a matrix printer or a PC/Laptop to Alpha RT com port. Select
service program #52SUP with the kV+ or kV- buttons. Enter the program
pressing the Autoexp button. Press Autoexp to exit program #52SUP.
Select service program #62Inf with the kV+ or kV- buttons. Press Autoexp
to enter the program. The info is printed out immediately. See an example
of the printout on next page.
The serial connection is an ordinary asynchronous receiver/transmitter
(RS-232). The serial port is initialized in Alpha RT as follows:
– 1200 baud if printer or Alpha ID II is connected
– 9600 baud if Dataview (or Alpha ID) is connected (in SETUP
parameter i =2)
– 8,N,1 (8 databits, No parity [always 0], 1 stop bit)
The effect of the 'no parity bit' and the stop bit is a total of two stop bits, so
receiving end must be configured correspondingly.
The data bytes are:
start
8-bit data 1 stop
The data can be received with a communications program, for example
Windows (Hyper)Terminal, to PC and printed afterwards on paper, or a
printer can be directly connected to the port.
48
GE Healthcare
32551-IMG rev 6