4 Alpha RT softstart procedure
Fig 4.9.
Preheat rises too slow
4.11
V-
K
REGULATION CHECK
1
Connect the DVM (averaging - NOT TRMS) on INVERTER BOARD to
TP22 and TP25 GND. Set the DVM to 0-10 Vdc range.
2
Set the unit to 28kV and 100 mAs (To get longer exposure time,
increase mAs. Make sure that the exposure is long enough for the
DVM reading to stabilize).
3
Make an exposure and check the voltage from the DVM during
exposure. It should be 2.8±0.01Vdc.
4
If not, adjust from trimmer potentiometer R63.
NOTE!
kV is set at the factory. For calibration use Unfors, RTI, PMX-I/M or similar
44
GE Healthcare
32551-IMG rev 6