Compression Force And Speed Adjustment With Ecs - GE Alpha RT 32551 Installation Instructions Manual

Table of Contents

Advertisement

5 Setups, alignments and adjustments
Fig 5.5.
C-arm display
5.2.7

Compression force and speed adjustment with ECS

NOTE!
The compression force is preadjusted at the factory and does not have to
be measured during the installation.
However, if the compression force needs to be changed, a bathroom
scale should be used. When using a bathroom scale, make sure that the
scale has soft material on top to simulate a more realistic compression.
The electrical compression limit can be adjusted on the C&Z DRIVER
33035-IMG (see additional adjusting instructions in ECS manual 32921-
IMG).
User (major) adjustment is done by turning the external switch (the cable
coming to connector X5) in the front of the system. Minor adjustment can
be done by adjusting trimpot R33 (clockwise to increase).
CAUTION!
When changing the maximum force, keep the external knob on maximum
position and when changing the minimum force, keep external knob on
minimum position.
Fig 5.6.
Adjustments on the C&Z Driver board 33035-IMG
68