5 Setups, alignments and adjustments
5.1.3

Service menu program #52 SUP

1
Enter the special program menu.
2
Select with kV+ button program number 52 (SUP).
3
Enter the program by pressing the AUTOEXP button.
4
Press kV+ and check following SUP subprograms:
#52SUP/A The density display shows A, which means the Auto Release
Time after each exposure in seconds.
1 Select a suitable time (normally 6 sec.) using the density-
and density+ buttons. Make sure that after an exposure
the Autorelease drives the paddle up to the upper limit.
2 Press kV+
#52SUP/L The density display shows L which means the Printer Label
Type for technique printout. Select using the density- and
density+ buttons. The types are:
Press kV+
#52SUP/H The density display shows H which indicates the Input Mode
of the text printed on the label. For each of the four rows
printed, the last 14 of 40 characters can be programmed.
See following table for how to program the label.
Sample of a label:
1234567890123456789012345678901234567890
B Rh 0.02s 28kV 002mAs +0 HOSPITALxNAMEx
6.5cm 200N 180^/R 0.06mGy CODExxxxxxxxxx
___/___/___
_________________________ DEPARTMENTxxxx
On row 1 the letter H in HOSPITAL is in character position 00,the letter O in
HOSPITAL is in position 01 etc. Remember that a "space" also consumes a
position.
The first letter in row 2 is C and the character position is 15 and continues
through position 29. Row 3 contains positions 30 through 44 and row 4
positions 45 through 59. See following the ASCII table and a Label
Programming Worksheet.
#52SUP/C A
additional compression force with pedal (ON/OFF)
B
set values (6) for compression force
50
0 = OFF
1 = all data in one row
2 = 36.7 x 101 mm label (U.S.A. default)
3 = 26.0 x 101 mm label
R1 12345 RADIOLOGISTxxx