Caldos27: Configuration and calibration of sample components

Configuring a new sample component
Calibrating a new sample component
EL3060 CONTINUOUS GAS ANALYZERS
In the configurator
In the configurator (see page 53), up to four sample components can be se-
lected from all the available sample components loaded in the gas analyzer in
the shortcut menu of the Caldos27 detector.
If necessary, an already configured sample component must be deleted. The
"Standard Gas" component configured ex works cannot be deleted. A "user
component" configured in the analyzer ex works can also not be deleted.
NOTE
If a sample component is deleted, its calibration data will be lost. It is there-
fore recommended that the ECT software tool is used to make a back-up
copy of the instrument data record before a sample component is deleted.
In the gas analyzer
In the gas analyzer, a sample component which is to be measured and dis-
played is selected from a maximum of five sample components in the
Active Component menu.
All available sample components are pre-calibrated ex works. The zero point
and end point must be checked and recalibrated if necessary after a user-
configured sample component has been activated for the first time.
NOTE
The sample components configured as per the order have been fully cali-
brated ex works.
|
OI/EL3060-EN REV. A