Password - ABB EL3060 Series Operating Instruction

Continuous gas analyzers
Table of Contents

Advertisement

102

Password

Password protection
Configuring the password
EL3060 CONTINUOUS GAS ANALYZERS
Access to the calibration and to those menus in which the configuration of
the instrument can be changed can be protected by a password. Password
protection is not activated ex works.
Password protection variants:
Access to the calibration menu is not included in the password protec-
tion.
Access to all instrument functions is included in the password protec-
tion (for instruments with SIL certification).
The password is set in the configurator (see page 53) in the "Options –
Password..." menu. It consists of a four-digit number; each of the numbers
may only have the values 1, 2 and 3 (example: "1213"). The setting "0000"
means that password protection is not activated.
|
OI/EL3060-EN REV. A